User’s Guide
7
AoFF - Automatic shutoff settings sub-menu
no Disabled
1 1-minute automatic shutoff
5 5-minute “ “
10 10-minute “ “
20 20-minute " "
30 30-minute " "
init - Initial (default) settings sub-menu
LBIN Pound-inch as default unit
KGMM Kilogram-millimeter as default unit
NCM Newton-centimeter as default unit
CCW Real time display at turn on
PEAK CW Peak clockwise display at turn on
PEAK CCW Peak counterclockwise display at turn on
CAL - Calibration sub-menu. See CALIBRATION section.
CALIBRATION
To properly calibrate the CTA, application of a precise torque value equal to the
full capacity of the tester in pound-inches (regardless of the displayed units) is
required.
While holding PEAK, turn on power to the tester. When ‘CAL’ appears on the
display, press ENTER three times to select the calibration mode. At the ‘null’
prompt, press ZERO. At the ‘SPAn’ prompt, apply the calibration torque and
press ENTER. The display will show ‘uuuu’ or ‘nnnn’ if the calibration torque is
insufficient or excessive, respectively. If this happens, the only way to terminate
the calibration mode is by momentarily disconnecting the battery or connecting
the AC adapter to the tester without plugging the other end into a wall outlet.
This will stop the calibration procedure without making any changes to the pre-
vious calibration data.
Successful calibration is indicated by ‘donE’ on the display. Press ENTER to
save the changes and resume normal operation.